12-9
5. If no information appears on the bottom of the NIBP parameter area, it indicates the airway
is in good condition and an air leak does not exist. However if the information
“PNEUMATIC LEAK” appears, it indicates the airway may have an air leak. In this case,
check for loose connections. After confirming all connections are secure, perform the test
again.
If there is still a failure, contact our Customer Service.
NOTE
z Set PAT TYPE to ADU in the PATIENT SETUP menu before leakage test.
z The pneumatic test, other than being specified in the EN 1060-1 standard, is to
be used to simply determine whether there are air leaks in the NIBP airway.
12.5 Maintenance and Cleaning
WARNING
z Do not press the rubber tube on the cuff with excessive strength.
z When a reusable cuff is not connected to the monitor or is being cleaned, always
cover the rubber tube with a lid. Avoid splashing liquid into the rubber tube or
the monitor inadvertently.
Reusable Blood Pressure Cuffs
The cuff is not suitable for dry-cleaned. Instead, it should be machine or hand washed. Hand
washing, may prolong the service life of the cuff. Before washing, remove the latex rubber
bladder. Allow the cuff to dry thoroughly after washing, and then reinsert the rubber bladder. The
cuff can be disinfected by means of conventional autoclaving, gas, or radiation disinfection in hot
air ovens, or sterilized by immersion in decontamination solutions. Remember to remove the
rubber bladder if you use this method