Parameter Menus Operations
2 - 20 0070-10-0666-01 Trio™ Operating Instructions
Modified Chest Lead (MCL) Monitoring
FIGURE 2-22 MCL Monitoring with a
3-wire Lead Set (AHA)
FIGURE 2-23 MCL Monitoring with a
3-wire Lead Set (IEC)
Place the RA (white) electrode under the
patient’s left clavicle, at the mid-
clavicular line within the rib cage frame.
Place the LA (black) electrode on the
right sternal border, at the fourth
intercostal space within the rib cage
frame.
Place the LL (red) electrode on the
patient’s lower left abdomen within the
rib cage frame.
Select ECG Lead I for MCL
1
monitoring.
Lead I is the direct electrical line between
the RA (white) electrode and the LA (black)
electrode.
Select ECG Lead II for MCL
6
monitoring.
Lead II is the direct electrical line between
the RA (white) electrode and the LL (red)
electrode.
Place the R (red) electrode under the
patient’s left clavicle, at the mid-
clavicular line within the rib cage frame.
Place the L (yellow) electrode on the
right sternal border, at the fourth
intercostal space within the rib cage
frame.
Place the F (green) electrode on the
patient’s lower left abdomen within the
rib cage frame.
Select ECG Lead I for MCL
1
monitoring.
Lead I is the direct electrical line between
the R (red) electrode and the L (yellow)
electrode.
Select ECG Lead II for MCL
6
monitoring.
Lead II is the direct electrical line between
the L (red) electrode and the F (green)
electrode.
LL
LA
RA
White
Red
Black
F
L
R
Red
Green
Yellow