  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 108
Trends Operations
2 - 80 0070-10-0666-01 Trio™ Operating Instructions
List Trend Menu Selections
1. INTERVAL
•To change the INTERVAL, select INTERVAL from the LIST TREND menu. (The
chosen interval can be saved as part of the “User Default Configuration” by
using the SAVE CURRENT function as described in section 2.3.5.9.)
Select from intervals of in 1min, 2min, 3min, 4min, 5min, 10min, 15min,
30min, or 1HR.
2. SCROLL
When an upward facing arrow appears on the upper portion of the menu, select
SCROLL and turn the knob clockwise to view more recent trend data.
When a downward facing arrow appears on the lower portion of the menu, select
SCROLL and turn the knob counterclockwise to view older trend data.
3. SCROLL
Select SCROLL to select the parameter group for which a view of trend data is desired.
Trend data for each parameter is displayed as follows:
•HR (bpm)
NIBP (Sys/Dia/Mean) (mmHg or kPa)
•SpO2
(%), PR (bpm)
IBP (Sys/Dia/Mean) (mmHg or kPa)
•RESP (rpm)
TEMP (°C or °F)
When a is displayed in the lower right corner of the menu, the user can scroll right
to view additional parameter data.
When a is displayed in the lower left corner of the menu, the user can scroll left to
view additional parameter data.
4. PRINT
To print the list trend data for all parameters, select PRINT from the LIST TREND
menu.
To stop the print job currently in progress, select PRINT on the front panel keypad.
Mark Event
The EVENT column of the LIST TREND menu displays time stamp event markers (A, B, C, or
D) that have been entered in the trend memory from the MARK EVENT menu (see section
2.3.4). The time stamp event markers are displayed in the EVENT column at the
corresponding trend entry time.
Remove List Trend Display
The LIST TREND menu/display does not automatically time-out. The user must manually exit
this menu. To exit the LIST TREND menu/display and return to the SYSTEM MENU, select
PREVIOUS MENU. To exit the LIST TREND menu/display and return to the normal screen,
select NORMAL SCREEN (from the menu or the Front Panel Keypad).