Trio™ Operating Instructions 0070-10-0666-01 4 - 5
User Maintenance Care and Cleaning of Accessories
NIBP cuffs can be sterilized with gamma sterilization without effecting the repeated
performance of the cuff. Steam sterilization is not recommended. Use of a washing liquid
containing bleach is not recommended because chlorine will chemically break down the
urethane on the inside of the cuff.
Antimicrobial Coating
Bladderless cuffs are treated with an antimicrobial coating. Antimicrobial technology
effectively controls a broad spectrum of bacteria, fungi, algae and yeasts on a wide variety
of treated substrates.
4.4.2.3 Disposable Blood Pressure Cuffs
Disposable cuffs are intended for single patient use only. Once a cuff is used on a patient it
should be discarded. Do not use the same cuff on any other patient. Do not sterilize or use an
autoclave on disposable cuffs.
NOTE: Disposable cuffs can be cleaned using a mild soap solution
and dried with a clean cloth. For cuffs with bladders,
remove bladder before cleaning.