Circuit Principle
2-7 Pulse oximeter Service Manual
2.1.1.10 Voltage Detect
DPM1 requires that the battery voltage and +7V be detected all the time. S1C33209 has 8
10-bit AD channels, two of which were used. S1C33209 cannot stand that high voltage, so,
a voltage divider was adopted to convert the monitored voltages to the range S1C33209
can detect. The divided voltage has a RC filter, which can resist the influence from
electrostatic charge.
2.1.2 Power Supply Circuit Block Diagram
Fig 2-10 Power Supply Circuit Block Diagram
2.2 Reliability Design
The following reliability control methods are used in the DPM1:
1. Derating for key elements;
2. Watchdog circuit, to restore the system when the software crashes;
3. By-pass capacitors for each IC, to reduce the interference from power supply;
4. Serial resistors on each line of the serial port, to avoid the destroy to devices due
to unintentional short with power;
5. One I/O port to control the write protection of the FLASH, to reduce the possibility
of unintentional FLASH write;
6. Monitoring the two major voltages through the CPU’s AD, to prevent inaccurate
measurement and unreliable data due to the voltage swing.