Circuit Principle
Pulse oximeter Service Manual 2-6
2.1.1.8 Serial Port Signal Voltage Level Conversion Circuit
MAX202E is used for the signal voltage level conversion. The output of MAX202E is 5V
TTL voltage, and the S1C33209’s I/O should use 74HC32 as the voltage level convertor
and driver and isolator. R116 and R115 convert the TTL level to LVTTL. R61 and R116 can
prevent short-circuit and electrostatic charge damage. Circuit block is shown as below:
Fig 2-8 Serial Signal Voltage Level Conversion Circuit
2.1.1.9 Button Circuit
DPM1 has three buttons, Power Button, ID Confirm Button and LCD Backlight Button.
The circuit is shown as below:
Fig 2-9 Button Circuit
R82 and C81 form RC filter circuit and can resist electrostatic charge interference. The
resistance of R82 is 4.7K, which can protect the circuit with big current. RC circuit can also
reduce the shaking influence.