Circuit Principle
2-5 Pulse oximeter Service Manual
R85 work together to prevent the locking of the CPU.
Fig 2-6 Delay Circuit Block
The power-on and power-off signals through two diodes constitute a AND logic output
signal named PCON signal, which directly controls the ON/OFF. Resistor R111 is
necessary for the AND logic circuit. R95, C82 and E12 form a RC filtering network to
protect the circuit from electrostatic charge.
Fig 2-7 Switch On Circuit