Mindray DPM 1 Manuel de service

Naviguer en ligne ou télécharger Manuel de service pour Détecteurs de monoxyde de carbone (CO) Mindray DPM 1. Mindray DPM 1 Service manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 206
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs

Résumé du contenu

Page 1 - Service Manual

DPM 6 Patient Monitor Service Manual

Page 2

4 6.3 Front housing Assembly... 6-3 6.3.1 12.1” LCD w

Page 3 - Revision History

5-20 5.3.15 Removing the Main Board Unplug all the cables on the mother board. The numbers beside the connectors indicates what device is connect

Page 4

5-21 2. Unscrew the 4 M3×6 screws and take out the main board assembly. NOTE z Since the main board assembly is connected with the power mod

Page 5 - Preface

5-22 5.3.16 Removing the Speaker 1. Unscrew the two M3×6 screws and remove the speaker. CPU board Butt socket to power module Button cell

Page 6 - Passwords

5-23 5.3.17 Removing the Power Module Assembly 1. Unscrew the four M3×12 screws and remove the power module assembly. 2. Unscrew the three

Page 7 - Contents

5-24 3. Lift the power board with a small flat screwdriver. Then, turn it over. 4. Unplug the cable between the AC input filter and the power

Page 8

5-25 5.3.18 Removing the Main Support 1. Unscrew the five M3×6 screws and disassemble the main support. Note z Be sure to remove the power

Page 9

5-26 2. Open the interface board cover and unscrew the four M3×6 screws, then, unplug the interface board assembly. 3. Unplug the cables on the

Page 10

5-27 4. Tear the insulating pad off the interface board. Unscrew the two M3×6 screws on the DVI interface board. Than unscrew the two screws besid

Page 11

5-28 5.4 Removing the SMR Assembly 1. First remove the 4 screw covers and then unscrew the 4 M3×8 screws. 2. Pull the left- and right-si

Page 12 - FOR YOUR NOTES

5-29 3. From the left side, remove the cable that connects the SMR interface board and the SMR communication board. Then take out the SMR from its

Page 13 - 1 Safety

5 7.3 Upgrading Functional Assemblies ... 7-5 7.3.1 Upgrading SMR ...

Page 14 - CAUTION

5-30 6. Release the clips and take out the Nios II module. Then unscrew the six M3×6 screws and remove the SMR communication board.

Page 15 - 1.2 Equipment Symbols

5-31 7. Use a socket wrench to unscrew the hexagon nut and countersunk external toothed lock washer assembly which can be further separated into t

Page 16

5-32 5.5 Disassembling Modules WARNING z For a reassembled module, a patient leakage current test must be performed before it is used again for

Page 17 - 2 Theory of Operation

5-33 2. Unlock the snap close by pressing it down about 1 mm with a flat screwdriver. At same time, push the snap plate forward with your thumb un

Page 18 - 2.2 System Connections

5-34 4. Remove the front panel carefully to avoid damaging the cables. The ICG module has an inner housing and an outer housing. Press the two clip

Page 19

5-35 6. Take out the non-isolated power board and disconnect all the cables from it. 7. Take out the infrared communication board and disconn

Page 20 - 2.3 Main Unit

5-36 5.5.2 Disassembling CO2 Module 1. Remove the four contact screws on the back and the two snap plates at the bottom and then unscrew the two

Page 21 - 2.3.1 Input System

5-37 3. Hold the inner housing with both hands. Press the snap side with your thumb and push the other side with the index finger and the middle

Page 22 - 2.3.2 Output System

5-38 4. Remove the module infrared communication backboard and disconnect all the cables from it.

Page 23 - Speaker

5-39 5. Disconnect the cables from the button board and unscrew the PT3×8 screw. Then release the two clips and take out the button board. After t

Page 24 - Main Board

6 FOR YOUR NOTES

Page 25 - Integral Module Rack

5-40 6. Unscrew the two screws on the front panel. Disconnect the tubing between the watertrap assembly and CO2 parameter board. Disconnect the ca

Page 26 - Power Module

5-41 7. Disconnect the tubing between the CO2 parameter board and the front panel. Then release the four clips and remove the CO2 parameter board.

Page 27

5-42 5.5.3 Disassembling the New MPM Module 1. Remove the contact screws and the spanner by referring to steps 1 to 3 as described in 5.5.1 Disass

Page 28 - USB_Hub board

5-43 3. Disconnect the cable of the infrared communication board, and release the snap lock to remove the infrared communication board. 4. Dis

Page 29 - CF Card Assembly

5-44 5. Unscrew the two M3×8 screws on the parameter board. Then pull off the parameter board rightward as shown below.

Page 30 - 2.4 Parameter Module

5-45 6. Unscrew the two M3×4 screws on the SpO2 board to separate the SpO2 board and the parameter board. 7. Unscrew the three M3×8 screws to s

Page 31 - 2.5 SMR

5-46 FOR YOUR NOTES

Page 32

6-1 6 Parts 6.1 Introduction This chapter contains the exploded views and parts lists of the main unit, satellite module rack and the parameter

Page 33 - 3 Testing and Maintenance

6-2 6.2 Main Unit 6.2.1 Exploded View 6.2.2 Parts List SN P/N Description Qty 6802-30-66761 or 6802-30-66762 Front housing assembly (with

Page 34 - 3.1.4 Recommended Frequency

6-3 6.3 Front housing Assembly 6.3.1 12.1” LCD with Anti-glare Screen Exploded View Parts List SN P/N Description Qty 1 043-000487-0

Page 35

1-1 1 Safety 1.1 Safety Information DANGER z Indicates an imminent hazard that, if not avoided, will result in death or serious injury. WARNI

Page 36 - 3.2.1 Visual Inspection

6-4 6 6802-30-66680 Power switch & LED board 1 7 115-004219-00 Button board 1 8 M04-003105--- Tapping screw PT3×8 3 9 M04-004012--- Cr

Page 37 - NIBP Accuracy Test

6-5 6.3.2 12.1” LCD with Touchscreen Exploded View Parts List SN P/N Description Qty 1 043-000487-01 or * 043-000490-01 Front bezel (DPM6

Page 38 - NIBP Leakage Test

6-6 9 M04-004012--- Crosshead screw M3×6 13 10 6802-30-66902 or 6802-30-66773(optional) 12.1" LCD assembly (touchscreen) 1 11 0000-10-10996

Page 39 - NIBP Calibration

6-7 6.3.3 12.1” Screen Assembly (with anti-glare screen) Exploded View

Page 40 - Module Tests

6-8 Parts List SN P/N Description Qty 1 0000-10-10752 or 0010-10-42633(optional) LCD TFT 12.1” 1 2 042-000784-00 or 042-000783-00 12” SHARP

Page 41 - Calibration

6-9 6.3.4 12.1” Screen Assembly (with touchscreen) Exploded View Parts List SN P/N Description Qty 1 0000-10-10752 or 0010-10-42633(optional) L

Page 42 - 3.2.4 AG Tests

6-10 6802-20-66650 AU screen cable 11 M04-004015--- Crosshead screw with washer M3×8 4 12 6802-20-66795 Screen cover 1 13 6802-20-66696 Touchsc

Page 43

6-11 Parts List SN P/N Description Qty 1 6802-30-66768 Fan assembly 1 2 6802-30-66765 Integral module rack 1 3 115-003229-00 Main suppor

Page 44

6-12 6.4.2 Battery Compartment Assembly Exploded View Parts List SN P/N Description Qty 1 6802-20-66714 Battery compartment 1 2 6802-30-6

Page 45

6-13 6.4.3 Power Module assembly Exploded View Parts List SN P/N Description Qty 1 6802-20-66719-51 Power supply compartment(mould MR66719)

Page 46 - 3.3 Power On Test

1-2 1.1.1 DANGER There are no dangers that refer to the product in general. Specific “Danger” statements may be given in the respective sections o

Page 47 - 3.4 Module Performance Tests

6-14 6.4.4 Interface Board Assembly 6.4.4.1 Interface Board Assembly (6802-30-66769) Parts List SN P/N Description Qty 1 042-000059-00 In

Page 48 - 3.4.4 NIBP Tests

6-15 6.4.4.2 Interface Board Assembly(6812-30-66994) Part List SN P/N Description Qty 1 042-000059-00 Interface board frame (new interfa

Page 49 - 3.4.6 IBP Tests

6-16 6.4.4.3 Interface Board Assembly(115-003230-00) Part List SN P/N Description Qty 1 042-000059-00 Interface board frame (new interface)

Page 50 - Method 2

6-17 6.4.5 Main Board Assembly Exploded View Parts List SN P/N Description Qty 1 M04-011001--- Nut, GB6170 M2.5 4 2 6800-30-50058 or *6812-3

Page 51 - 3.4.7 C.O. Test

6-18 6.4.6 Integral module rack Exploded View Parts List SN P/N Description Qty 1 6802-20-66709 Integral module rack 1 2 6802-20-66817 Water

Page 52 - 3.4.8 Mainstream CO

6-19 12 M04-060026--- Stud screw M3×7+6-6 coated with antirust nickel 5 13 6802-20-66710 Grounding leaf 2 14 M04-004012--- Crosshead screw M3

Page 53 - 3.4.10 AG Tests

6-20 6.4.8 Rear Housing Assembly Exploded View Parts List SN P/N Description Qty 1 6802-20-66699 Handle 1 2 6802-20-66737 Waterproof stri

Page 54 - 3.4.12 BIS Test

6-21 6.4.9 CF Card Assembly CF Card Assembly (115-001906-00) Parts List SN P/N Description Qty 1 6802-20-66725 Card cover 1 2 6802-20-66726

Page 55 - 3.4.13 RM Test

6-22 CF Card Slot Cover Kit(6812-30-66995) Parts List SN P/N Description Qty 1 047-000731-00 Cover overlay 1 2 6802-20-66725 Card cover 1

Page 56 - 3.4.14 CCO/SvO

6-23 6.4.10 6802 Internal Wireless AP Assembly Exploded View Parts List SN P/N Description Qty 1 045-000196-00 Chip radiator 1 2 0000-10-10777

Page 57

1-3 1.1.4 Notes NOTE z Refer to Operation Manual for detailed operation and other information. 1.2 Equipment Symbols Attention: Consult accom

Page 58 - 3.9 Recorder Check

6-24 6.4.11 Recorder Assembly Exploded View Parts List SN P/N Description Qty 1 TR6F-20-67300 Recorder chamber 1 2 M04-004012--- Crosshea

Page 59 - 3.10 Network Print Test

6-25 6.5 SMR 6.5.1 SMR Assembly Exploded View Parts List SN P/N Description Qty 1 6800-20-50249-51 SMR inside assembly 1 2 M6G-020015---

Page 60 - 3.11 Battery Check

6-26 6.5.2 SMR Inside Assembly Exploded View Parts List SN P/N Description Qty 1 6800-20-50519 Insulating plate 1 2 9201-20-36010 Torsion spri

Page 61 - 3.12 Factory Maintenance

6-27 16 M04-004012--- Crosshead screw M3×6 6 17 M04-000306--- Stud screw M3×10+8-8 coated with antirust nickel 1 18 6800-30-50075 Nios II mo

Page 62 - 3.12.4 Software Version

6-28 8 6800-20-50275 2-slot module inner housing, right side 1 9 M51A-30-80924 NIBP&MPM analog assembly 1 10 M51A-30-80870 MPM front housi

Page 63 - 3.12.6 Calibrate NIBP

6-29 Parts List SN P/N Description Qty 1 M90-000156--- Plastic hex nut M3 2 051-000976-00 M51A Multi-parameter module, 5-lead, standard 051-

Page 64 - Maintenance and Test Report

6-30 SN P/N Description Qty 17 801-6800-00211-00 NIBP pump for M51A and iPM 1 18 S1-0103-00-0511 Tee connector, White, Nylon 3 19 0010-10-12

Page 65

6-31 Parts List SN P/N Description Qty 1 6800-20-50264-54 Single-slot module front panel (C.O./English) 1 2 6800-20-50284 Key pad for sing

Page 66

6-32 6.6.4 RM Module Exploded View Parts List SN P/N Description Qty 1 6800-20-50264-53 Single-slot module front panel (RM/English) 1 2

Page 67

6-33 13 6800-30-50812 RM module infrared communication board 1 14 M04A-30-58452 RM module PCB assembly 1 15 6800-20-50270 Single-slot module

Page 68

1-4 FOR YOUR NOTES

Page 69 - 4 Troubleshooting

6-34 Parts List SN P/N Description Qty 1 0010-10-42826 ICG module signal wire 1 2 6800-20-50266-51 Single-slot module front panel (ICG/Eng

Page 70 - 4.6 Troubleshooting Guide

6-35 6.6.6 AG Module Exploded View Parts List SN P/N Description Qty 1 6800-20-50276-51 AG front panel (English) 1 2 6200-20-11614 Gas o

Page 71 - 4.6.2 Display Failures

6-36 6 6800-20-50394 latch 1 7 6800-21-50310 Bisx connecter cable 1 8 6800-20-50164 Module fan and cable 1 9 6800-20-50540 Module fan cushion

Page 72 - 4.6.3 Module Rack Failures

6-37 6.6.7 BIS Module Exploded View

Page 73

6-38 Parts List SN P/N Description Qty 1 6800-20-50264-51 Single module front housing(BIS/English) 1 2 6800-21-50310 BIS connector cable 1 3

Page 74 - 4.6.4 Alarm Problems

6-39 6.6.8 IBP Module Exploded View Parts List SN P/N Description Qty 1 6800-20-50265-51 Single-slot module front panel(IBP/English) 1 2 68

Page 75 - 4.6.6 Recorder Failures

6-40 14 6800-20-50268 Single-slot module rear housing 1 15 6800-20-50338 Waterproof seal 01 1 16 6800-30-50097 Module infrared communication b

Page 76 - 4.6.8 CF Card Problems

6-41 Parts List SN P/N Description Qty 1 6800-20-50275 2-slot module inner housing, right side 1 2 M04-003105--- Tapping screw PT3×8 1 3

Page 77

6-42 6.6.10 Oridion CO2 Module Exploded View Parts List SN P/N Description Qty 1 6200-20-11614 Outlet 1 2 9201-20-35914 Microstream CO2 co

Page 78

6-43 15 6800-20-50278 Snap plate 1 16 6800-20-50338 Waterproof seal 01 1 17 6800-20-50292 Fixer 2 18 6800-30-50806 Oridion CO2 module infrared

Page 79

2-1 2 Theory of Operation 2.1 Introduction This patient monitor is designed to monitor a fixed set of physiological parameters including ECG, hea

Page 80

6-44 Parts List SN P/N Description Qty 1 043-000473-00 CCO/SvO2 module front panel (English) 1 2 6800-20-50284 Silicon button 1 3 6800-20-50282

Page 81 - 5 Repair and Disassembly

6-45 6.7 Remote Display Box Exploded View Part List SN P/N Description Qty 1 M12A-20-75103 CIS lock catch 1 2 M12A-20-75107 Lock catch spring

Page 82 - 5.3 Disassembling Procedure

6-46 16 M04-004012--- Cross pan head screw with washer M3×6 10 17 6800-30-50058 MCF547x CPU module (basic configuration/lead-free process) 1 18

Page 83

6-47 5 9201-20-36009-51 or 9201-20-35917-51(configured with Dlink AP) Wireless network card installation board (ASUS) Wireless network card installa

Page 84

6-48 6800-30-51132 Encoder assembly 801-6800-00145-00 Encoder assembly 1 801-6802-00010-00 Power switch & LED board 1 009-000247-00 Con

Page 85

6-49 801-6800-00004-00 MPM module rack communication board 1 801-6800-00005-00 NiosII module 1 6802-20-66799 Cable connecting the main unit

Page 86

6-50 6.9.3 New MPM Material P/N Description Qty 0671-00-0102-01 Nellcor SpO2 board (MDU) 1 051-000943-00 9008 V2.0 SpO2 board 1 040-001149

Page 87

6-51 configuration/Mindray algorithm/single heart rate meter), FRU 801-M51A-00012-00 New M51A integrative parameter board (5-lead/standard configura

Page 88

6-52 6.9.4 Parameter Modules P/N description Qty 6800-30-50486 BIS Module 1 6800-30-50488 RM module 1 6800-30-50407 CAPNOSTAT CO2 module 1

Page 89

6-53 115-010761-00 New MPM-7 module (Mindray SpO2/3/5 lead/without IBP) 1 115-013388-00 New MPM-8 module(Masimo 2013 SpO2/3/5 lead/without IBP) 1 11

Page 90 - 5.3.7 Removing the Inverter

2-2 2.2 System Connections 2.2.1 Mounting the Patient Monitor The patient monitor can be mounted on a wall bracket or on a trolley support. The w

Page 92

7-1 7 Upgrade 7.1 Introduction You can upgrade parameter modules, functional assemblies and software by connecting the patient monitor to a PC ru

Page 93

7-2 7.2 Upgrading Parameter Modules You can upgrade the following parameter modules: Parameter module PN Description Remark M51A-30-80873 MPM-1m

Page 94

7-3 115-013388-00 New MPM-8 module(Masimo 2013 SpO2/3/5 lead/without IBP) 1 115-010750-00 New MPM-9 module(Nellcor SpO2/3/5 lead/without IBP) 1 115-

Page 95

7-4 115-001698-00 AG module upgrade package (with O2, without accessories) / 115-013203-00 AG module upgrade package (with BIS, without O2 and acce

Page 96

7-5 7.3 Upgrading Functional Assemblies You can upgrade the following functional assemblies: Functional assembly PN Description Remark 6800-30

Page 97

7-6 7.3.3 Upgrading Recorder Refer to corresponding section to install the recorder in your patient monitor. Refer to the Operator’s Manual for t

Page 98

7-7 7.4 Upgrading Software You can upgrade the software with the System Update Tool (PN: 6800-30-50771) through network.. This tool can directly

Page 99

7-8 110-001839-00 M51A V2.0 DSP(BF512) software 110-001978-00 M51A V2.0 DSP(Mortara) software program 110-001842-00 9008 V2.0 SpO2 software 7.4.1

Page 100

7-9 7.4.1.1 Upgrading BIOS program If your BIOS program is version 1.2 or prior, the patient monitor has a blank screen during software upgrade. Fo

Page 101

2-3 2.2.2 Connectors for Peripheral Devices On the back of the patient monitor you will find all connectors for peripheral devices. 1. AC Power

Page 102 - 5.3.16 Removing the Speaker

7-10 7.4.1.3 Upgrading FPGA Program If your FPGA program is version 1.0, first you have to upgrade it to version 2.0 using the System Update Tool v

Page 103

A-1 A Electrical Safety Inspection The following electrical safety tests are recommended as part of a comprehensive preventive maintenance program.

Page 104

A-2 A.2 Device Enclosure and Accessories A.2.1 Visual Inspection Test Item Acceptance Criteria No physical damage to the enclosure and accessori

Page 105

A-3 Preparation 1. First select the test current that will be used for performing the Protective Earth Resistance test by pressing AMPERES (SOFT K

Page 106

A-4 NOTE z When "Over" is displayed for Ohms, this signifies that a valid measurement was not obtained because either an open connection

Page 107

A-5  SOFT KEY 1 toggles the DUT outlet Polarity from Normal to Off to Reverse.  SOFT KEY 2 toggles the DUT outlet from Earth to No Earth.  SO

Page 108

A-6 A.6 Patient Leakage Current Patient leakage currents are measured between a selected applied part and mains earth. All measurements have a true

Page 109

A-7 In Case of Failure  Check any broken of the enclosure. Replace any defective part.  Inspect wiring for bad crimps, poor connections, or dam

Page 110

A-8 Preparation To perform a calibration from the Mains on Applied Part test, press CAL (SOFT KEY 2). 1. Disconnect ALL patient leads, test lead

Page 111

A-9 NOTE z If all of the applied parts correspond to the instrument type, the applied parts will be tied together and one reading will be taken. I

Page 113

2-4 2.3 Main Unit The patient monitor consists of:  Input system: button board, knob, touchscreen, power switch and LED board  Output system:

Page 114

A-10 5. Press SOFT KEYS 1-4 to select leakage tests 6. Press APPLIED PART (SOFT KEY 4) at any time to select the desired applied part leakage cur

Page 115

A-11 A.9 Scheduled Electrical Safety Inspection For scheduled electrical safety inspection, test items 1, 2, 3, 4, 5, 6, 7, and 8 included in the

Page 116 - Module

A-12 A.10 Electrical Safety Inspection after Repair The following table specifies test items to be performed after the equipment is repaired. Repa

Page 117

A-13 ELECTRICAL SAFETY INSPECTION FORM Location: Technician: Equipment: Control Number: Manufacturer: Model: SN: Measurement equipment /SN: Date

Page 118

A-14 FOR YOUR NOTES

Page 120

Mindray DS USA, Inc.800 MacArthur Blvd.Mahwah, New Jersey 07430 USATel:1.800.288.2121Tel:1.201.955.8000www.mindra

Page 121

2-5 2.3.1 Input System Button board The button board, located at the lower part of the monitor’s front panel, contains 6 keys and provides connect

Page 122

2-6 2.3.2 Output System LCD The patient monitor adopts a high-resolution LCD. The LCD is connected with the main board. The signals and power su

Page 123

2-7 Module Description Power interface Introduces a DC from the main board. Recorder power module Converts the input power into voltages that fit

Page 124

2-8 2.3.3 Processing and Communications System Main Board The main board is the heart of the patient monitor. It implements a series of tasks incl

Page 125

2-9 The mother board is in charge of connections and communications with other components and provides the following interfaces: Name Descriptio

Page 126

2-10 2.3.4 Power Management System Battery The patient monitor uses two chargeable lithium-ion batteries (11.1 V, 4500 mAh). The battery compartme

Page 127 - 6 Parts

2-11 The following diagram shows the pins of the power socket connecting the power module and the mother board: Pin ID Marking Description 1/3/

Page 128 - 6.2 Main Unit

2-12 2.3.5 Equipment Interface System USB_Hub board The USB_Hub board is connected with the mother board. It is compatible with USB1.1 connector

Page 129 - 6.3 Front housing Assembly

2-13 DVI Interface Board The DVI interface board is connected with the mother board. The following diagram shows its interfaces to other compone

Page 130 - Encoder mounting board

I Intellectual Property Statement Mindray DS USA, Inc. (hereinafter called Mindray DS) owns the intellectual property rights to this product and

Page 131 - Parts List

2-14 2.4 Parameter Module Each parameter module may consist of the module infrared communication board, module power board, module button board, p

Page 132 - Spring

2-15 2.5 SMR The satellite module rack (SMR) is independent of the patient monitor. It provides 8 slots for mounting parameter modules. It has the

Page 133

2-16 FOR YOUR NOTES

Page 134

3-1 3 Testing and Maintenance 3.1 Introduction To ensure the patient monitor always functions normally, qualified service personnel should perform

Page 135

3-2 3.1.2 Test Report Upon completion of the tests, the table of preventative maintenance test reports and the table of maintenance test reports in

Page 136 - 6.4 Main Unit

3-3 SpO2 test / Pressure check Leakage test NIBP test and calibration Calibration Temp test / Performance test IBP test and calibration Pressure ca

Page 137

3-4 Electrical safety tests Refer to A Electrical Safety Inspection. 1. Following any repair or replacement 2. After the monitor drops. 3. At least o

Page 138

3-5 3.2.2 NIBP Tests and Calibration NIBP Accuracy Test Tools required:  T-shape connector  Appropriate tubing  Balloon pump  Rigid Vessel w

Page 139 - 6.4.3 Power Module assembly

3-6 NOTE z You can use an NIBP simulator to replace the balloon pump and the reference manometer to perform the test. z You can use an appropriate

Page 140

3-7 2. Raise the pressure in the rigid vessel to 250 mmHg with the balloon pump. Then, wait for 5 seconds to let the measured values becoming stable

Page 141 - Part List

II FOR YOUR NOTES

Page 142

3-8 3.2.3 Sidestream and Microstream CO2 Module Tests Leakage test Follow this procedure to perform the test: 1. Plug the module into the module ra

Page 143 - 6.4.5 Main Board Assembly

3-9 5. Open the relief valve to vent standard CO2 and make sure that there is an excess gas flow through the T-shape connector to air.. 6. Check

Page 144 - 6.4.6 Integral module rack

3-10 3.2.4 AG Tests Leakage Test 1. Plug the AG module into the module rack. 2. Wait for a minute until the AG module warmup is finished and then

Page 145 - 6.4.7 Main Support Assembly

3-11 5. Adjust the relief valve and make sure the flowmeter reading is stable and within 10 and 50 L/min. 6. Check that the concentration of each c

Page 146 - 6.4.8 Rear Housing Assembly

3-12 4. Open the relief valve and vent a certain standard gas or gas mixture. Then adjust the relief valve and make sure the flowmeter reading is

Page 147 - 6.4.9 CF Card Assembly

3-13 3.2.5 Preventative maintenance test report Customer name Customer address Servicing person Servicing company Equipment under test (EUT)

Page 148

3-14 Block the gas inlet of the module or watertrap. The sidestream CO2 flowrate is slower than 10ml/min and an alarm of CO2 Filterline Err is given.

Page 149

3-15 3.4 Module Performance Tests 3.4.1 ECG Tests and Calibration ECG Performance Test Tool required:  Fluke Medsim 300B patient simulator recomme

Page 150 - 6.4.11 Recorder Assembly

3-16 3.4.2 Resp Performance Test Tool required:  Fluke Medsim 300B patient simulator recommended Follow this procedure to perform the test: 1. Co

Page 151 - 6.5 SMR

3-17 3.4.5 Temp Test Tool required:  Resistance box (with accuracy above 0.1Ω) Follow this procedure to perform the test: 1. Connect the two pins

Page 152 - 6.5.2 SMR Inside Assembly

III Preface Manual Purpose This manual provides detailed information about the assembling, dissembling, testing and troubleshooting of the equip

Page 153 - 6.6 Parameter Modules

3-18 IBP Pressure Calibration Method 1 Tools required:  Medsim300B patient simulator, MPS450, or other equivalent device IBP adapter cable for tes

Page 154 - 6.6.2 New MPM Module

3-19 5. Adjust the calibration value in the [Maintain IBP] menu until it is equal to the reading of sphygmomanometer 6. Select the [Calibrate] but

Page 155

3-20 3.4.8 Mainstream CO2 Tests NOTE z Select [Main Menu]→[Maintenance >>]→ [User Maintenance >>]→enter the required password→[Maintain

Page 156 - 6.6.3 C.O. Module

3-21 In the figure above, 1 A steel gas cylinder with 6±0.05% CO2 2 Flowmeter 3 3-way valve (power supply controlled) 4 Open to air 5 Power

Page 157

3-22 3.4.11 ICG Test Tool required:  ICG simulator (BZ-4575)  BioZ Dx patient cable (PN: 040-000543-00)  BioZ Dx lead wire (PN: 040-000544-00)

Page 158 - 6.6.4 RM Module

3-23 Method 2: Tools required:  BIS simulator (with flexible cable)  BIS cable (PN: 6800-30-50761) 1. Connect the BIS sensor with the BIS simul

Page 159 - 6.6.5 ICG Module

3-24 5. Select [Respiratory Loop] in the [RM Setup] menu. Verify that the displayed TV is within 500±50ml and RR is within 20±1rpm. 3.4.14 CCO/SvO2

Page 160

3-25 3.5 Nurse Call Relay Performance Test Tools required:  Oscilloscope 1. Connect the nurse call cable to the Nurse Call Connector of the pati

Page 161 - 6.6.6 AG Module

3-26 3.7 Electrical Safety Test See A Electrical Safety Inspection for electrical safety tests. 3.8 Touchscreen Calibration Tools required:  Non

Page 162

3-27 3.10 Network Print Test Note z HP LaserJet 1505n or 2035n laser printer is recommended for BeneView series of patient monitors. Tools require

Page 163 - 6.6.7 BIS Module

IV Abbreviations Abbreviations used in this manual are: MPM multi-parameter module SMR satellite module rack CMS central monitoring system Pass

Page 164

3-28 3.11 Battery Check Tools required:  None. Function Test 1. If the patient monitor is installed with batteries, remove the batteries first.

Page 165 - 6.6.8 IBP Module

3-29 3.12 Factory Maintenance 3.12.1 Accessing Factory Maintenance Menu To access the factory maintenance menu, select [Main Menu]→ [Maintenance >

Page 166 - 6.6.9 Mindray CO

3-30 3.12.4 Software Version Selecting [Software Version] will show software version information. The [Software Versio n] menu is as follows:

Page 167

3-31 3.12.5 Monitor Information Selecting [Monitor Information] will show the status of the patient monitor. Monitor information is displayed as fol

Page 168 - 6.6.10 Oridion CO

3-32 Maintenance and Test Report (See the above sections for detailed test procedures and contents) Customer name Customer address Servicing per

Page 169 - 6.6.11 CCO Module

3-33 ECG waves are displayed correctly without noise and the HR value is within 80±1 bpm. Yes N o ECG Lead Off alarm behaves correctly. Yes

Page 170

3-34 Block the gas inlet of the module or watertrap. The sidestream CO2 flowrate is slower than 10ml/min and an alarm of CO2 Filterline Err is given.

Page 171 - 6.7 Remote Display Box

3-35 respectively. Nurse call relay performance test The relay contacts are close when an alarm occurs. Yes No Analog output performance test Th

Page 172 - 6.8 Wireless AP

3-36 FOR YOUR NOTES

Page 173 - 6.9 Replaceable Parts

4-1 4 Troubleshooting 4.1 Introduction In this chapter, patient monitor problems are listed along with possible causes and recommended correctiv

Page 174

1 Contents 1 Safety ...

Page 175 - 6.9.2 SMR

4-2 4.4 Software Version Check Some troubleshooting tasks may require you to identify the configuration and software version of your patient monit

Page 176 - 6.9.3 New MPM Material

4-3 4.6.2 Display Failures Symptoms Possible Cause Corrective Action Cables defective or poorly connected. 1. Check that the cable from the disp

Page 177

4-4 Cables defective or poorly connected. 1. Check that the cables from the touchscreen to the touchscreen control board, the touchscreen control bo

Page 178 - 6.9.4 Parameter Modules

4-5 Power supply failure 1. Check if the voltage between two contact screws in any slot reaches 12V DC. If yes and the parameter module functions p

Page 179

4-6 3-slot or MPM module rack communication board failure Replace the 3-slot or MPM module rack communication board. Nios II module failure Repla

Page 180

4-7 4.6.5 Button and Knob Failures Symptoms Possible Cause Corrective Action Cable defective or poorly connected 1. Check that cable between but

Page 181 - 7 Upgrade

4-8 4.6.7 Output Interface Failures Symptoms Possible Cause Corrective Action Respective output disabled 1. Select [Main Menu]→[Analog Output Se

Page 182

4-9 4.6.9 Power Supply Failures Symptoms Possible Cause Corrective Action Battery failure Replace battery. Cable defective or poorly connected

Page 183

4-10 NOTE z When the power module has a failure, it may cause problems to other components, e.g. the monitor suddenly breaks down during start-up,

Page 184

4-11 4.6.11 Software Upgrade Problems Symptoms Possible Cause Corrective Action Bootstrap upgrade fails Power failure or unintended power off du

Page 185 - 7.3.1 Upgrading SMR

2 3.3 Power On Test ...3-14 3.4 Module P

Page 187 - 7.4 Upgrading Software

5-1 5 Repair and Disassembly 5.1 Tools During disassembly and replacing, the following tools may be required:  Phillips screwdrivers  Small fla

Page 188

5-2 5.3 Disassembling Procedure 5.3.1 Removing the Recorder 1. Open the recorder door and unscrew the two M3×6 screws. 2. Pull the two clip

Page 189

5-3 3. Unscrew the M3×6 screw and unplug the recorder grounding cable and the cable between the recorder and the mother board. 4. Pull the t

Page 190

5-4 5. Pull the press bar upwards about 1 mm and then unplug the flexible cable. Remove the cable that connects the driving board and the button b

Page 191 - Electrical Safety Inspection

5-5 7. Unscrew the PT2×6 screw. Prize the thermal printhead a little using a small flat-bladed screwdriver to remove the printhead.

Page 192 - A.3 Device Labeling

5-6 5.3.2 Separating the Front and Rear Housing 1. Remove the hook and than unscrew the four M3×12 screws, one of which can be seen only when t

Page 193

5-7 NOTE z Exercise care when pulling the front housing out. Be sure not to damage the cables and connectors. z Avoid pressing the knob on the ta

Page 194 - A.5 Earth Leakage Test

5-8 5.3.3 Removing the Power Switch & LED Board 1. Unplug the connecting cable between power switch & LED board and the button board. U

Page 195 - In Case of Failure

5-9 5.3.5 Removing the Button Board 1. Disconnect the cables from the button board to the alarm LED board, backlight board and touchscreen cont

Page 196 - A.6 Patient Leakage Current

3 4.6.5 Button and Knob Failures ... 4-7 4.6.6 Recorder Failures ...

Page 197

5-10 5.3.6 Removing the Touchscreen Control Board 1. Unplug the cables respectively from the touchscreen and button board to the touchscreen con

Page 198

5-11 2. Unplug the cables respectively from the button board and the LCD to the backlight board. Then, unscrew the two M3×6 screws to remove the i

Page 199 - Preparation

5-12 2. Unscrew the four M3×6 screws underneath the screen cover and then remove the LCD. 5.3.9 Removing the Alarm LED Board 1. After removin

Page 200

5-13 5.3.10 Removing the Fan Assembly Unplug the cable that connects the fan assembly and the mother board. Then, unscrew the four M3×6 screws and

Page 201

5-14 5.3.12 Removing the Integral Module Rack 1. Disconnect the cable that connects the integral module rack and the mother board. Unscrew

Page 202

5-15 2. Disconnect the cable that connects the MPM module rack communication board and the 3-slot module rack communication board. Release the tw

Page 203

5-16 4. Unscrew the hex nut assy using the sleeve. Then separate the washer, spring and contact screw from each other. 3-slot module rack c

Page 204

5-17 5.3.13 Removing the CF Card Assembly 1. Unplug the cable that connects the CF card assembly and the mother board. Then, unscrew the two M3×6

Page 205

5-18 5.3.14 Removing the wireless AP assembly 5.3.14.1 Removing the internal wireless AP 1. Disconnect the wireless AP cable. Then unscrew the t

Page 206

5-19 5.3.14.2 Removing the external wireless AP 1. Disconnect the network cable and the power cord 2. Press the leaf on the fix board with for

Commentaires sur ces manuels

Pas de commentaire